CB-7021, CB-7022, CB-7024 Users Manual
CB-7021/21P Voltage Output Calibration Sequence:
1. Connect voltmeter to module’s voltage output.
2. Warm-Up for 30 minutes.
3. Set type to 32. (0 to 10V) Refer Sec. 2.1.
4. Program output for 10V. Refer Sec. 2.7.
5. Check Refer Sec. 2.13.