Excitation Source
Channels: 1
Signal Connection: BNC
Frequency Range: 1 Hz to 5 kHz
Frequency Resolution: 0.01 Hz
Amplitude Settings (p-p): 10V, 5V, 2V, 1V, 500mV, 200mV, 100mV, 0mV
Waveform Modes: Continuous sine, Sweep sine, Custom sine
Output Impedance: 50 Ohm
Accuracy: ±0.1dB
DC Accuracy, Excluding Noise
Applies to 2-Pole and Bypass Filter Modes
Accuracy at 0° to 50°C (32° to 122°F)
± % of Reading
Range
Typical Maximum
Offset
±25V
0.5 1 ±15mV
±5V
0.15 0.3 ±3mV
±2.5V
0.15 0.3 ±2mV
±1V
0.15 0.3 ±1mV
±500mV
0.15 0.3 ±1mV
±250mV
0.15 0.3 ±1mV
±100mV
0.15 0.4 ±1mV
±50mV
0.15 0.4 ±1mV
±25mV
0.2 0.4 ±1mV
These numbers are valid for 1 year after calibration and over the
entire operating temperature range of the unit.