Section 3 Command Descriptions
3.23
M8
will generate a
Service Request
when the
Digital488
self-test fails.
Refer to the
Test
command
(T0)
description for details on self-tests.
M16
will generate a
Service Request
when the
Digital488
has completed the
execution of a set of commands from the bus controller. This is used by the
controller to assure the completion of a set of commands before sending a
subsequent set of commands.
Example:
CLEAR 718
reset the
Digital488
OUTPUT 718;"M4X"
select SRQ on Bus error
OUTPUT 718;"F7X"
send an invalid bus command.
Note: ERROR and SRQ LEDs should illuminate
CLEAR 718
reset the
Digital488
OUTPUT 718;"M1XM4X"
select SRQ on Bus error and SRQ on Service input.
OUTPUT 718;"M5X"
This has the same effect as the command above
where
M1X
plus
M4X
equals
M5X
.